Multi 3400i
Operation
37
4.4.4
AutoRead AR (Drift control)
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable effect on the repro-
ducibility of the measured value.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing
only change to another measuring mode after completion of AutoRead.
1
Call up the
or SAL measuring mode with
▲ >
<
▼>.
2
Immerse the conductivity measuring cell in the test sample.
3
Activate the AutoRead function with
sured value is frozen (hold function).
4
Start AutoRead with
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
5
If necessary, start the next AutoRead measurement with
6
To terminate AutoRead: Press the