4 operation, 1 switching on the measuring instrument, Operation – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 15: Switching on the measuring instrument, 4operation

background image

Multi 3400i

Operation

15

4

Operation

4.1

Switching on the measuring instrument

Measuring mode when

switching on

Note
The measuring instrument has an energy saving feature to avoid un-
necessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour.

The energy saving feature is not active

z if the power is supplied by the plug-in power supply,
z if the AutoStore function is active,
z if the communication cable and a PC with a running communication

program are connected,

z if the printer cable is connected (for external printers).

AutoRange measure-
ment range selection

There are several measuring ranges available for both oxygen and
conductivity measurements. As a rule, the measuring instrument is in
the measuring range with the highest possible resolution. If a measur-
ing range is exceeded, AutoRange causes the measuring instrument to
change automatically to the next higher measuring range.
The AutoRange function is always active with oxygen and conductivity
measurements and cannot be switched off.

Note
The measuring ranges for oxygen and conductivity measurements can
be found in chapter 7 T

ECHNICAL

DATA

.

1

Press the key.
The display test appears briefly on the display.
After this, the measuring instrument automatically switches to
the measuring mode.

Connected sensor

Measuring mode

No sensor or pH/ORP electrode

pH or ORP measurement (de-
pending on the last selected set-
ting)

D. O. probe or conductivity mea-
suring cell

Last selected measuring mode

2 sensors of any type

Last selected measuring mode