Data storage

pH 7310

48

ba75929e03

12/2011

Settings

You can configure the automatic storage function with the following settings:

Starting the auto-

matic storage func-

tion

To start the automatic storage function, select Continue with <><> and
confirm with . The meter switches to the measured value display.

The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura-
tion.

Energy saving mode

([Eco mode])

If the automatic storing function is active, the meter provides an energy saving
mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
saving mode switches off functions of the meter that are not required for the

Menu item

Possible set-
ting

Description

ID number

1 ... 10000

ID number for the dataset series.

Interval

1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min

Storage interval.

The lower limit of the storage
interval can be restricted by the
number of free storage locations.
The upper limit is restricted by the
storage duration.

Duration

1 min ... x min

Storage duration.
Specifies after which time the
automatic storage should be ter-
minated.

The lower limit of the storage
duration is restricted by the stor-
age interval.
The upper limit is restricted by the
number of free storage locations.

1 Remaining storage duration
2 Graphical display of the storage duration

If the automatic storage function is activated, only the following
keys are active: Softkeys, , and . The other
keys and the automatic switch-off function are deactivated.

1

2

pH

6.949

25.0

°C

0d03h00min

01.07.2011

08:00

0

14

USB output

Eco mode