2 buffer sets for calibration, Buffer sets for calibration – Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual
Page 39

pH 7310
Settings
ba75929e03
12/2011
39
7.1.2
Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
Calibration
/Buffer
TEC
NIST/DIN
ConCal
...
Buffer sets to be used for pH cali-
bration (see section 5.2 C
Calibration
/One
point calibration
Yes
No
Quick calibration with 1 buffer
Calibration
/Cali-
bration interval
1 ... 7 ... 999 d
Calibration interval
for the pH elec-
trode (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring window.
Calibration
/Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst
slope of -59.2 mV/pH (100 x deter-
mined slope/Nernst slope).
Man. temperature
-25 ... +25 ...
+130 °C
Entry of the manually determined
temperature. For measurements
without temperature sensor only.
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahr-
enheit.
All temperatures are displayed
with the selected unit.
Resolution pH
0.001
0.01
0.1
Resolution of the pH display
Stability control
On
Off
Switches on or off the automatic
stability control during measure-
ment (see section 5.1.1 M
Reset
-
Resets all sensor settings to the
delivery condition (see section
7.4.1 R
Menu item
Possible set-
ting
Description