Appendix: Turbidity values under 1 FNU/NTU

With turbidity values (under 1 FNU/NTU), the measured value is
strongly influenced by the cell and its alignment.

In order to increase measuring accuracy with turbidity values under 1
FNU/NTU, calibration in the 0.02 FNU/NTU standard and later
measurement should be carried out in the same cell. For calibration in
the standards 10.0 and 1000 FNU/NTU follow the instructions on the
display.

Proceed as follows to measure turbidity values under 1 FU/NTU:

Calibration procedure:

Calibrate

! after the calibration interval has expired
! with a temperature change

Measuring

1

Press the key.

The menu-guided calibration begins.

2

Carry out calibration of the standards 1000 FNU/NTU and
10.0 FNU/NTU in a clean and unscratched cell according to
the menu guidance.

3

Fill the cleaned cell with the 0.02 FNU/NTU standard and
calibrate.

4

Mark the alignment of the cell.

5

Fill the marked and cleaned cell with test sample, align it
with the marking and measure.

6

If necessary, fill the marked and cleaned cell once more
with test sample and carry out further measurements.