Calibration record

At the end of each calibration procedure a calibration info (

i

symbol)

and the calibration record is displayed.

Displaying and

downloading calibration

data to interface

You can view the data of the last calibration on the display.
Subsequently, you can download the displayed calibration data to the
interface, e. g. to a printer or PC, with the key.

The calibration record of the last calibration can be found under the
menu item, Configuration / Turbidity / Calibration record.

Sample printout of a

record

Preparing the

calibration

Perform the following preparatory activities when you want to calibrate:

Note
For turbidity values under 1 FNU/NTU, follow the instructions given in
the appendix (see A

PPENDIX

: T

URBIDITY

VALUES

UNDER

1 FNU/NTU).

Carrying out calibration

31.10.03 16:13
Turb 430 IR Ser. no. 12345678
Calibration Turbidity
Calibration date

31.10.03 16:13:33

Calibration interval

90 d

1

Keep the cells with the required calibration standards ready
and mark them as necessary (see page 33).

2

Clean the cell (see section 5.2.2).

3

Insert the cell (see section 4.2).

1

Press the key.

The menu-guided calibration begins.
Follow the instructions on the display.

2

Insert the cell with the displayed calibration standard (here e.g.
1000 FNU/NTU) in the cell shaft (see section 4.2).

i Insert standard

1000 FNU/NTU

i Press and hold
i Align sample

Turbidity \ Calibration