2 turbidity, 1 turb430ir, 2 turb430t – Xylem TURB 430 PORTABLE TURBIDITY METER User Manual

Page 59: Turbidity 7.2.1, Turb 430 ir, Turb 430 t

background image

Turb 430 IR/T

Technical data

59

ba75507e03

07/2006

7.2

Turbidity

7.2.1

Turb 430 IR

7.2.2

Turb 430 T

Measuring principle

Nephelometric measurement according to DIN EN ISO 7027

Light source

Infrared LED

Measuring range

0.01 ... 1100 FNU/NTU

Resolution

in the range
0,01 ... 9,99

max. 0.01 FNU/NTU

in the range
10,0 ... 99,9

max. 0.1 FNU/NTU

in the range
100 ... 1100

max. 1 FNU/NTU

Accuracy

in the range
0 ... 1100 FNU/NTU

± 2 % of the measured value
or ± 0.01 FNU/NTU

Reproducibility

0,5% of the measured value

Response time

4 seconds

Calibration

Automatic 3-point calibration

Measuring principle

Nephelometric measurement according to US EPA 180.1

Light source

White light tungsten lamp

Measuring range

0.01 ... 1100 NTU

Resolution

in the range
0,01 ... 9,99

max. 0.01 NTU

in the range
10,0 ... 99,9

max. 0.1 NTU

in the range
100 ... 1100

max. 1 NTU

Accuracy

in the range
0 ... 500 NTU

± 2 % of the measured value
or ± 0.01 NTU

in the range
500 ... 1100 NTU

± 3 % of the measured value

Reproducibility

1% of the measured value

Response time

7 seconds

Calibration

Automatic 3-point calibration