6 saving, 1 storing measurement datsets, Saving 4.6.1 – Xylem TURB 430 PORTABLE TURBIDITY METER User Manual
Page 40: Storing measurement datsets

Operation
Turb 430 IR/T
40
ba75507e03
07/2006
4.6
Saving
The meter has 1000 memory locations for measurement datsets.
You can transmit measured values (datasets) to the data memory with
the
Each storage process transmits the current dataset to the interface at
the same time.
The number of memory locations that are still free is displayed in the
Store menu. The number of memory locations that are occupied is
displayed in the System Measured value memory menu.
Measurement dataset
A complete dataset consists of:
! Date/time
! ID number (ID)
! Measured value
4.6.1
Storing measurement datsets
Proceed as follows to transmit to the data memory and simultaneously
output to the interface a measurement dataset:
Note
A measurement dataset is stored quickly by twice pressing
is stored with the ID last set.
1
Press the
The Store display appears.
2
Using <
▲> <▼>,
change and confirm the ID number (ID) as necessary (0 ...
999).
3
Using
The dataset is stored. The instrument switches to the
measured value display.
i 02.02.2005 11:24:16
16.80 FNU/NTU
ID:
1
Store (ID: 1)
Store
(
996 free
)