1 connecting a pc/external printer, 2 configuring the rs232 interface, Connecting a pc/external printer – Xylem TURB 430 PORTABLE TURBIDITY METER User Manual
Page 44: Configuring the rs232 interface

Operation
Turb 430 IR/T
44
ba75507e03
07/2006
4.7.1
Connecting a PC/external printer
Connect the interface to the devices via the AK540/B (PC) or AK540/S
(external printer) cable.
Caution
The RS232 interface is not galvanically isolated. When connecting
an earthed PC/printer, measurements cannot be performed in
earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Socket assignment
4.7.2
Configuring the RS232 interface
For error-free data transmission, the RS232 interface should be set to
the same transmission speed (Baud rate) on the Turb 430 IR/T and
PC/printer.
You can set the following values for the baud rate on the Turb 430 IR/
T: 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu, Configuration / System /
Interface / Baud rate.
Baud rate
can be selected from: 1200, 2400, 4800, 9600,
19200
The baud rate must agree with the baud rate set on
the PC/printer.
Handshake
RTS/CTS
PC only:
Parity
none
Data bits
8
Stop bits
1s
RS 232
REC
1
3
4
5
6
2
1 RTS
2 RxD
3 TxD
4 DTR
5 SG
6 CTS