Maintenance, cleaning, disposal
53
ba75507e03
07/2006
5.2
Cleaning
Occasionally wipe the outside of the meter with a damp, lint-free cloth.
Disinfect the housing with isopropanol as required.
Caution
The housing components are made out of synthetic materials
(polyurethane, ABS and PMMA). Thus, avoid contact with acetone
and similar detergents that contain solvents. Remove any
splashes immediately.
5.2.1
Cleaning the cell shaft
If liquid is in the cell shaft (e.g. due to a spilled cell), clean the cell shaft
as follows:
5.2.2
Cleaning the cells
Cells have to be clean, dry, and free of fingerprints and scratches.
Therefore, clean them regularly:
Note
Scratches in the glass change the optical characteristics of the cell and
falsify the measured value. For this reason, never use scratched cells!
1
Switch the Turb 430 IR/T off and pull out the power plug.
2
Rinse the cell shaft with distilled water.
1
Clean the cells inside and out with hydrochloric acid or
laboratory soap.
2
Rinse out several times with distilled water.
3
Let them dry in the air.
4
Only hold the cells by the top or by the light protection cap so
that the optical path is not impaired.
5
Before measuring, clean the cell with the enclosed cleaning
cloth.