beautypg.com

LaMotte COD Plus Colorimeter 1922-EX2 User Manual

Page 92

background image

PROCEDURE

1.

Press and hold

ON

button until colorimeter turns on.

2.

Press

ENTER

to start.

3.

Press

ENTER

to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 14 Fluoride)
from TESTING MENU.

5.

Scroll to and select 14 Fluoride from menu.

6.

This test requires a reagent blank. Rinse a clean tube (0290) with clear,
colorless, fluoride free water. Fill to the 10 mL line with clear, colorless,
fluoride free water.

7.

Use the 0.5 mL pipet (0353) to add 0.5 mL of *Sodium Arsenite Solution
(4128). Cap and mix.

8.

Use the 1.0 mL pipet (0354) to add 2 measures of *Acid-Zirconyl
SPADNS Reagent (3875). Cap and mix thoroughly. (This is the reagent
blank.)

9.

Insert tube into chamber, close lid and select SCAN BLANK.

10. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with

sample water. Repeat steps 7 and 8.

11. Insert tube into chamber, close lid and select SCAN SAMPLE. Record

result.

12. Press

OFF button to turn colorimeter off or press EXIT button to exit to a

previous menu or make another menu selection.

COD PLUS TEST PROCEDURES 6/02

FLUORIDE 3/3

This manual is related to the following products: