FOR EXAMPLE:
If a fresh water sample at 20

°C has a pH of 8.5 and the test result is 1.0 ppm as

Total Ammonia-Nitrogen:
1.

The percentage from the table is 11.18% (or 0.1118).

2.

1 ppm Total Ammonia-Nitrogen x 0.1118 = 0.1118 ppm Unionized
Ammonia-Nitrogen.

3.

Total Ammonia-Nitrogen

1.0000 ppm

Unionized Ammonia-Nitrogen

-

0.1118 ppm

Ionized Ammonia-Nitrogen

=

0.8882 ppm