INTERFERENCES: The following substances produce a positive interference at

the concentration given:
Chloride (Cl–)

7000 mg/L

Phosphate (PO

4

–3

)

16 mg/L

Hexametaphophate (NaPO

3

)

6

1 mg/L

The following substances produce a negative interference at
the concentration given:
Alkalinity (CaCO

3

)

5000 mg/L

Aluminum (Al

3+

)

0.1 mg/L

Iron (Fe

3+

)

10 mg/L

Sulfate (SO

4

–2

)

200 mg/L

Color and turbidity must be removed or compensated for in
the procedure. Temperature should be maintained within
5°C of room temperature.