Storing – LaMotte COD Plus Colorimeter 1922-EX2 User Manual
Page 31
A reverse font (dark background with light
characters) will appear to indicate that the reading
can be adjusted. Use
s or t to scroll to the
concentration of the sample, 0.10 ppm in this
example.
002 Alumninum
0.09
* Calibrate
Set the calibration by pressing
Q/ENTER to select
Calibrate
.
002 Aluminum
0.10
* Calibrate
Two menu choices will be offered, Set
Calibration
and Factory Setting. Set the
calibration by pressing
Q/ENTER to select Set
Calibration
; or use
s or t to scroll to and select
Factory Setting
to revert to the factory
calibration.
002 Aluminum
0.10
* Set Calibration
Facctory Setting
The meter will display the message “Storing” and
return to 002 Aluminum test.
Storing
The calibration for 002 Aluminum has now been
standardized and can be used for testing. The
standardization can be removed by repeating the
calibration and selecting Factory Setting.
002 Aluminum
* Scan Sample
Next Test
Previous Test
%/Abs
Calibrate
Scan Blank
COD PLUS COLORIMETER 4/03
31