beautypg.com

Blank done, Storing – LaMotte COD Plus Colorimeter 1922-EX2 User Manual

Page 43

background image

Place a clear blank in the sample chamber. Press the

Q/ENTER button to scan the blank.

Insert Blank

* Continue

The screen will display Blank Done for about 1
second.

Blank Done

* Scan Blank

Place the reacted low standard in the sample
chamber. Press

Q/ENTER to scan the low standard.

Insert Lo Standard

* Continue

Place the reacted high standard in the sample
chamber. Press

Q/ENTER to scan the high

standard.

Insert Hi Standard

* Continue

The meter will display the message “Storing” and
return to the EDIT USER TEST01 menu.

Storing

EDIT USER TEST01

* New Calibration

Decimal Places

Select Units

COD PLUS COLORIMETER 6/02

43

This manual is related to the following products: