Procedure - fresh water – LaMotte COD Plus Colorimeter 1922-EX2 User Manual
Page 69
PROCEDURE - FRESH WATER
1.
Press and hold
ON
button until colorimeter turns on.
2.
Press
ENTER
to start.
3.
Press
ENTER
to select TESTING MENU.
4.
Select ALL TESTS (or another sequence containing
4 Ammonia-NLF
) from TESTING MENU.
5.
Scroll to and select 4 Ammonia-NLF from menu.
6.
Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.
7.
Insert tube into chamber, close lid and select SCAN BLANK
.(See Note.)
8.
Remove tube from colorimeter. Use the 1.0 mL plastic pipet (0354) to add
2.0 mL of *Salicylate Ammonia #1 (3978). Cap and mix.
9.
Use the 0.15 g spoon (0727) to add two measures of *Salicylate #2
Reagent (7457). Cap and mix until dissolved. Wait 1 minute.
10. At end of 1 minute waiting period use 0.1 g spoon (0699) to add two
measures of *Salicylate #3 Reagent Powder (7458). Cap and shake
vigorously for at least 30 seconds and all solid has dissolved. Wait 12
minutes for maximum color development.
11. At the end of the 12 minute waiting period, immediately mix and insert
tube into chamber, close lid and select SCAN SAMPLE. Record result.
12. Press
OFF
button to turn colorimeter off or press
EXIT button to exit to a
previous menu or make another menu selection.
CALCULATIONS:
To express results as Unionized Ammonia (NH
3
):
ppm Unionized Ammonia (NH
3
) =
ppm Ammonia-Nitrogen (NH
3
–N) x 1.2
To express results as Ionized Ammonia (NH
4
):
ppm Ionized Ammonia (NH
4
+
) =
ppm Ammonia-Nitrogen (NH
3
–N) x 1.3
To determine the percentages of Unionized and Ionized Ammonia-Nitrogen,
consult the Appendix.
NOTE: For the best possible results, a reagent blank should be determined to
account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents are obtained.
AMMONIA-NITROGEN - HIGH RANGE 2/2
COD PLUS TEST PROCEDURES 2/04