YSI ProODO User Manual
Page 10
10
11
An example of a GLP record
(Operation Performed is % DO Calibration)
*** Calibrate – DO ***
Date:
11/01/2008 MM/DD/YY
Time:
04:03:05 PM
User ID:
Tech0001
Probe ID:
SN: 08D
User Field #1
Cloudy
Sensor:
08F000015
Method:
DO Air Calibrate
Cal Value:
100.0 %
Sensor Value:
-0.591150
Salinity Mode:
Manual
Salinity Value:
0.000000 SAL PSU
Barometer
767.09 mmHg
Temperature
19.2 °C
Calibrate Status:
Calibrated
_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
In the System menu, highlight GLP and press
enter to view and modify the GLP settings.
Highlight Options and press enter to access
User ID, Probe ID, User Defined Fields, and
Re-Cal Prompt.
User ID may be used to identify the person
calibrating the instrument. Highlight User ID
and press enter to select or edit a User ID from
a list of previously entered IDs. Or, highlight
Add New and press enter to create a new User
ID using the alpha/numeric entry screen. The
User ID may also be changed in the Calibration
menu during the calibration process. The
selected User ID will be stored in the GLP file
with each calibration record. A User ID could
be a person’s initials or badge number. The
character limit is 16 characters.
Probe ID is stored with the calibration record
and may be used to distinguish one cable/
probe assembly from another, typically by serial number. Highlight Include
Probe ID and press enter to turn this function on and off. There will be a check
mark in the box when the function is enabled. Highlight Probe ID and press
enter to add, view, edit, or select a Probe ID. Probe ID may also be changed in
the Calibration menu during the calibration process. The character limit is 16
characters.
User Fields 1 and 2 are stored with the calibration record and may be used
to enter other parameters pertinent to the user, such as weather conditions,
elevation, etc. Highlight Include User Field 1 or Include User Field 2 and press
enter to turn this function on and off. Highlight User Field 1 or User Field 2
and press enter to add, view, edit, or select a User Field. The character limit is
16 characters. When enabled, a prompt for selecting a User Defined Field will
appear during the calibration process.
Re-Cal Prompt may be used to remind the user
to perform a calibration. To set a time interval,
highlight DO and press enter to access the
numeric entry screen. Enter a value in days and
press enter to confirm the reminder time. To
turn off the Re-cal prompt, set the reminder to
zero (0) days (this is the default).
The Security section of the GLP menu is a password protected area. This area
includes options to set a new password and to lock access to the calibration
menu. When first viewing the security menu, you will be required to enter a
password. Use the “shift” on the alpha/numeric screen to switch to lower case
and enter “ysi123”. This is the default password.
Protect Cal can be enabled or disabled. When enabled, the user must know and
enter the instrument’s password to enter the calibration menu option. Highlight
Protect Cal and press enter to enable (
Q
) or disable (
P
) this feature.
Set Password allows a user to set the security password. Highlight Set Password
and press enter. Using the alpha/numeric entry screen, enter the new password.
The password can have up to 16 characters.
Contact YSI Technical Support at [email protected] or +1 937 767-7241 if
you forget or misplace your password.
Once a password is set, and the GLP security screen
exited, a password must be entered to make changes
under GLP security. Keep passwords in a safe place.