YSI Professional Plus User Manual
Page 7

4
5
1010 dual cable
Port 1 Options
Port 2 Options
pH
pH
oRP
oRP
pH or pH/oRP*
pH or pH/oRP*
ammonium
ammonium
chloride
chloride
nitrate
nitrate
none (port plug)
* If using a 6051010 cable, a sensor must be installed in port 1 for correct
operation. If you install a pH/oRP combo sensor into a 6051010 cable, oRP
will not be measured. It is not recommended to use a pH/oRP combo sensor
on a 6051010 cable.
1020 dual cable
Port 1 Options
Port 2 Options
pH
Polarographic Do
oRP
Galvanic Do
pH or pH/oRP
none (port plug)
ammonium
chloride
nitrate
none (port plug)
If using a 1020 cable, install a pH, oRP, pH/oRP, Ammonium, Nitrate or
Chloride sensor in port 1 and a Do sensor in port 2.
If using a 605103 pH/ORP combination probe on a
6051020 or 6051030 cable you can report both pH
and ORP. However, it is recommended to set ISE1 as
pH and ISE2 as ORP in the Sensor Setup menu.
the Quatro cable bulkheads are labeled
1, 2, Do, and Ct, see figure to the left. All
sensors except the Conductivity/temperature
sensor can be installed following the Standard
Pro Series Sensor Installation instructions.
Conductivity/temperature sensor installation
is described in the next section. For ease of
installation, YSI recommends that you install
a sensor into port 1 first; followed by Do
installation, then port 2, and lastly C/t.
Quatro
Cable (pn:
605790)
Port 1
Options
Port 2
Options
DO Port
Options
CT Port Options
pH
pH
Polarographic
Do
5560
Conductivity/
temperature
sensor only
oRP
oRP
Galvanic Do
pH or pH/
oRP*
pH or pH/
oRP*
none (port
plug)
ammonium
ammonium
chloride
chloride
nitrate
nitrate
none (port
plug)
* If using a Quatro cable, a sensor must be installed in port 1 for correct
operation of port 2. If you install a pH/oRP combo sensor into a Quatro
cable, oRP will not be measured. It is not recommended to use a pH/oRP
combo sensor on a Quatro cable.
Before installing either dissolved oxygen sensor, the instrument
must be configured for the sensor being installed. See
the Setup - Dissolved Oxygen section of this manual for
instrument configuration instructions. Failure to do this
may result in damage not covered under warranty.
First, ensure both the sensor connector and sensor port on the cable are
clean and dry. to connect the sensor, grasp the sensor with one hand and the
sensor connection end of the cable (bulkhead) in the other. Push the sensor
into the connector on the cable until it is properly seated and only one o-ring
is visible. Failure to properly seat the probe may result in damage. twist the
sensor clockwise to engage threads and finger tighten (Figure 2). Do not use a
tool. this connection is waterproof. Please refer to the sensor installation sheet
that is included with each sensor for detailed instructions.