YSI Professional Plus User Manual
Page 19
28
29
After selecting your User ID and/or Probe ID if appropriate, highlight Do and
press enter.
Highlight DO % and press enter to confirm.
the instrument will use the internal barometer
during calibration and will display this value
in brackets at the top of the display. Highlight
Barometer and press enter to adjust it if needed.
If the barometer reading is incorrect, it is
recommended that you calibrate the barometer.
Note - the barometer should be reading “true”
barometric pressure (see Barometer section
for more information on “true” barometric
pressure). If the value is acceptable, there is
no need to change it or perform a barometer
calibration.
the Salinity value displayed near the top of the
screen is either the salinity correction value
entered in the Sensor menu or the Salinity
value as measured by the conductivity sensor
in use and enabled. If you are not using a
conductivity sensor, the Salinity correction
value should be the salinity of the water you
will be testing. Highlight Salinity and press
enter to modify this setting if necessary. See the
Salinity Correction section of this manual for
more information.
Wait for the temperature and Do% values under “Actual Readings” to stabilize,
then highlight Accept Calibration and press enter to calibrate. or, press Esc
to cancel the calibration. If User Field 1 or 2 are enabled in the GLP menu, you
will be prompted to select these inputs and then press Cal
to complete the
calibration. the message line at the bottom of the screen will display “Calibrating
Channel...” and then “Saving Configuration...”.
Calibrating Do% in Water Saturated Air:
2-Point Calibration with Zero Solution
Place the sensor in a solution of zero Do.
A zero Do solution can be made by dissolving approximately
8 - 10 grams
of
sodium sulfite (Na
2
So
3
) into 500 mL tap water or DI water. Mix the solution
thoroughly. It may take the solution 60 minutes to be oxygen-free.
Press Cal
. Highlight Probe ID or User ID if you wish to add, select, edit, or
delete an ID. Probe ID must be enabled in the System GLP menu to appear in
the Calibrate menu. When enabled, these IDs are stored with each calibration
record in the GLP file.
After selecting the Probe ID and/or User ID if appropriate, highlight Do and
press enter. Highlight Zero and press enter. Wait for the temperature and
Do% values under “Actual Readings” to stabilize, then press enter to Accept
Calibration. If User Field 1 or 2 are enabled, you will be prompted to select the
fields and then press Cal
to complete the calibration. the screen will then
prompt for a follow-up second point calibration.
Highlight DO% and press enter to continue with the next calibration point.
Rinse the sensor of any zero oxygen solution using clean water. then follow the
steps under Calibrating Do % in Water Saturated Air to complete the second
point.
Calibrating in mg/L or ppm as a titration:
1-Point Calibration
Place the sensor into an adequately stirred sample that has been titrated to
determine the dissolved oxygen concentration. Allow the sensor to stabilize.
Press Cal
. Highlight Probe ID or User ID if you wish to add, select, edit, or
delete an ID. Probe ID must be enabled in the System GLP menu to appear in
the Calibrate menu. When enabled, these IDs are stored with each calibration
record in the GLP file.
After selecting the Probe ID and/or User ID if appropriate, highlight Do and
press enter. Highlight DO mg/L or ppm and press enter.