3 electrical connections, 2 radio transmission, 3 instrument safety – YSI MIQ/Blue PS Set User Manual
Page 41: Electrical connections -3, Radio transmission -3, Instrument safety -3, 2 radio transmission 6.3 instrument safety, Awg 24 0.2 mm
MIQ/Blue PS
Technical data
6 - 3
ba76095e02
01/2013
6.1.3
Electrical connections
Terminal strip inside the
enclosure
Terminals
Cable glands
6.2
Radio transmission
6.3
Instrument safety
Used directives and
standards
In addition to the standards listed in the IQ S
ENSOR
N
ET
system operat-
ing manual, the MIQ/Blue PS is based on the following directives and
standards:
X13
X12
100...
240V AC
NETZ/MAINS
L1
N
X11
X10
24V DC
EINGANG
INPUT
POWER
SENSORNET 2
SENSORNET 1
ON
EIN
OFF
AUS
X6
RO
T
RED
RO
T
RED
SN
TERMINA
TO
R
SCHIRM
SHIELD
GRÜN
GREEN
SCHIRM
SHIELD
GRÜN
GREEN
X5
X4
X3
X2
X1
SENSORNET 3
X8
X9
X7
RO
T
RED
SCHIRM
SHIELD
GRÜN
GREEN
+
-
Terminal type
Screw-type terminal strip, accessible by
opening the lid
Terminal ranges
Solid wires:
Flexible wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
0.2 ... 2.5 mm
2
Suitable for cable diam-
eter
4.5 ... 10 mm or 7 ... 13 mm
Radio frequency
2.402 ... 2.480 GHz
Transmitting power
max. 20 dBm or 100 mW
Bridgeable distance
max. 100 m (328 ft)
The specified maximum range applies if the
radio link was installed and tested according
to this operating manual (see section 3.3).
Data radio
– EEC directive 1999/5/EC
(R&TTE directive)
– EN 300 328-2