6 connecting the voltage supply, 1 connection to 100 v ac, Connecting the voltage supply -16 3.6.1 – YSI MIQ/Blue PS Set User Manual
Page 26: Connection to 100 v ac -16
Installation
MIQ/Blue PS
3 - 16
ba76095e02
01/2013
3.6
Connecting the voltage supply
The following two sections need only be observed if the IQ S
ENSOR
N
ET
island should be supplied with power by the MIQ/Blue PS.
3.6.1
Connection to 100 ... 240 V AC
Warning
If the power supply is connected incorrectly, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
The MIQ/Blue PS may only be connected by a trained electri-
cian.
The connection of the MIQ/Blue PS to the power supply may
only be carried out when it is not carrying any voltage.
The power supply must fulfill the specifications given on the
nameplate and in chapter 6 T
.
When installed in a building, a switch or power switch must be
provided as an interrupt facility for the MIQ/Blue PS.
The interrupt facility must
– be installed in the vicinity of the MIQ/Blue PS, easily acces-
sible by the user, and
– be labeled as an interrupt facility for the MIQ/Blue PS.
After it has been installed, the MIQ/Blue PS may only be opened
if the mains voltage has been switched off beforehand.
Materials required
Wire end sleeves, suitable for the power line, with suitable crimping
tool
1 x cable gland, suitable for the cable diameter (see section 3.5 on
Tools
Cable stripping knife
Wire stripper
Phillips screw driver
Small screw driver.