background image

MIQ/Blue PS

Installation

3 - 15

ba76095e02

01/2013

General installation

instructions

Observe the following points when attaching connecting wires to the
terminal strip:

 Shorten all the wires to be used to the length required for the instal-

lation

 Always fit all the ends of the wires with wire end sleeves before con-

necting them to the terminal strip

 Any wires that are not used and project into the enclosure must be

cut off as closely as possible to the cable gland.

 In each of the remaining free openings, screw in a small cable gland

with sealing ring and close them with a blind plug.

Warning

No free wires must be allowed to project into the enclosure. Oth-
erwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. This could result in life
threatening electric shock when working with the IQ S

ENSOR

N

ET

.

Always cut off any wires that are not in use as closely as possible
to the cable gland.