Page 30

Installation
MIQ/Blue PS
3 - 20
ba76095e02
01/2013
3.6.2
Connection to 24 V DC
Warning
If the 24 V DC supply is connected incorrectly, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
The MIQ/Blue PS may only be connected by a trained electri-
cian.
The 24 V DC supply must meet the specifications quoted on the
nameplate and in chapter 6 T
(protective low volt-
age SELV).
The connection of the MIQ/Blue PS to the power supply may
only be carried out when it is not carrying any voltage.
When installed in a building, a switch or power switch must be
provided as an interrupt facility for the MIQ/Blue PS.
The interrupt facility must
– be installed in the vicinity of the MIQ/Blue PS, easily acces-
sible by the user, and
– be labeled as an interrupt facility for the MIQ/Blue PS.
Note
Battery systems should be protected against total discharge. The MIQ/
Blue PS has no integrated deep discharge protection.
Materials required
Wire end sleeves, suitable for the 24 V AC/DC feed line, with suit-
able crimping tool
1 x cable gland, suitable for the cable diameter (see section 3.5 on
Tools
Cable stripping knife
Wire stripper
Phillips screw driver
Small screw driver.
Preparing the 24 V DC
line
1
Cut off the cable to the required length.
2
Strip the cable insulation for approx. 45 mm.
3
Bare the wires 1 and 2 and fit them with wire end sleeves.