8 technical data, 1 measurement characteristics, Technical data -1 – YSI NiCaVis 705 IQ User Manual
Page 49: 1 measurement characteristics -1, Chapter 8 t, Echnical, Data, Ee chapter 8 t, 8technical data

NiCaVis 705 IQ
Technical data
8 - 1
ba75950e01
03/2013
8
Technical data
8.1
Measurement characteristics
Measuring principle
Spectrophotometric absorption measurement; integrated micropro-
cessor electronics, shielded 2-wire connection for
power and data transmission.
Wavelength range
200 - 720 nm
Measuring gap (optical
layer thickness)
Nitrate measurement,
5 mm gap width
Carbon measurement,
5 mm gap width
The quoted measuring ranges are nominal measuring ranges that are
theoretically possible. In practice, real measuring ranges exist that are
given by the limits of photometric determination. The limits are signifi-
cantly influenced by the light scattering due to solids and the absorption
of accompanying substances (sample matrix). Therefore, no fixed
measuring range can be determined beforehand.
NiCaVis 705 IQ
5 mm
Measuring
location
Measuring
mode
Measuring range
Resolution
Outlet
NO3
0.0 ... 250.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 50.00 mg/L
0.01 mg/L
Measuring
location
Measuring mode
Measuring range
Resolution
Outlet
COD spectr., total
0.0 ... 800.0 mg/L
0.1 mg/L
COD spectr., dis
-
solv
0.0 ... 800.0 mg/L
0.1 mg/L
TOC spectr., correl.
0.0 ... 500.0 mg/L
0.1 mg/L
DOC spectr., correl.
0.0 ... 500.0 mg/L
0.1 mg/L
BOD spectr., correl.
0.0 ... 500.0 mg/L
0.1 mg/L
SAC 254, total
0.0 ... 600.0 1/m
0.1 1/m
SAC 254, dissolved
0.0 ... 600.0 1/m
0.1 1/m
UVT 254
0.0 ... 100.0 %
0.1 %