YSI NiCaVis 705 IQ User Manual
Page 21

NiCaVis 705 IQ
Commissioning
3 - 11
ba75950e01
03/2013
Response time t90
10 ... 60 min
Response time of the signal smoothing.
Depending on the sample matrix, mea-
sured values can vary more or less
strongly (e.g. due to foreign bodies or
air bubbles).
The signal filter (here, the
Response
time t90
)
reduces the limits of variation
of the measured values.
The
Response time t90
specifies the
time after which 90 % of a signal
change are displayed.
This menu item is only available with:
Signal smoothing
=
On
The specified time is used as a guide
value for low (10 min) to high (60 min)
signal smoothing and can deviate from
the value that was set.
Cleaning mode
On
Off
Switching on/off the automatic, time-
controlled cleaning function.
Cleaning type
Ultrasound
Air
Ultrasound+air
Cleaning methods
This menu item is only available with:
Cleaning mode
=
On
The duration of the cleaning procedure
is defined in the sensor for all cleaning
types.
Note: The duration of the cleaning pro-
cedure with air can be adjusted for the
linked relay in the menu,
150 - Settings
of outputs and links
/
Relay function
/
Sensorgesteuert
/
Pulsdauer
.
Please note the software requirements
for the MIQ/CR3, MIQ/R6, MIQ/CHVP-
lus, DIQ/S 182 relays (see section 3.1).
Save and quit
-
The system confirms the saving of the
settings and the display switches to the
next higher level.
Quit
-
The display switches to the next higher
level without saving the new settings.
Menu item
Settings
Explanations