What to do if...

NiCaVis 705 IQ

7 - 2

ba75950e01

03/2013

Drifting measured values

Display of

OFL

Measured values

fluctuating heavily

Cause

Remedy

– Cleaning efficiency too low -

measurement windows

becoming more and more

contaminated

– Clean the measurement win-

dows (section 5.2.4)

– Use other cleaning method
– Select different measurement

location

– Set longer cleaning duration
– Set more frequent cleaning

procedures

– Install optional compressed air

cleaning device

– Failure of the optional com-

pressed air cleaning system

– Clean the measurement win-

dows (section 5.2.4)

– Check all components of the

compressed air cleaning sys-

tem (compressor, compressed

air line, UV/VIS sensor, rinsing

openings)

– Substantial change of test

sample (sample matrix)

– Carry out new user calibration

Cause

Remedy

– Measuring range exceeded

– Check process

– Foreign matter in the measur-

ing gap

– Clean the measuring gap (sec-

tion 5.2.3)

– Erroneous user calibration

– Repeat user calibration. Make

sure the samples for compari-

son are representative.

Cause

Remedy

– Gas bubbles in front of the

measurement windows

– Check the position within the

process and the installation

position