3 sensor check/zero adjustment, Sensor check, See section 4.3.3) – YSI NiCaVis 705 IQ User Manual
Page 32
Measuring / Operation
NiCaVis 705 IQ
4 - 8
ba75950e01
03/2013
4.3.3 Sensor check/
Zero adjustment
Note
The sensor check or Zero adjustment must be carried out under abso-
lutely clean conditions. If you do not work carefully enough the Zero ad-
justment can deteriorate the measuring quality.
Ultrapure water
(H2O dist.)
As ultrapure water exclusively use distilled or deionized water that is
suitable for analysis purposes. Especially make sure that the concen-
tration of metrologically relevant substances (organic carbon
compounds, nitrate) is negligible.
Storage location of the
Zero adjustment
data
The data of the sensor checks are not stored.
The data of the user zero adjustment are stored in the sensor.
Steps
A successful
Sensor check (H2O dest.)
is the prerequisite for a valid
user calibration. After each
Sensor check (H2O dest.)
the sensor is
evaluated.
A changed sensor evaluation proves that the cleaning was effective.
The steps of the sensor check or
Zero adjustment
are identical except
for the used check solutions (ultrapure water or standard solution) and
are described together below. The sensor check and
Zero adjustment
are carried out with the aid of the sensor sleeve.