Measuring / Operation

NiCaVis 705 IQ

4 - 6

ba75950e01

03/2013

Performing a user

calibration

Proceed as follows if you want to perform a user calibration:

In the

Settings of sensors and diff. sensors

menu, set the

Signal

smoothing

setting to

Off

.

Subsequently, leave the menu with

Save and quit

.

How to navigate to the

Settings of sensors and diff. sensors

menu and

the operation of the menu is described in section 3.4.3.

At least one value pair and a maximum of two value pairs can be

entered for a valid calibration. The values must be in ascending order.

If they are not in ascending order, a calibration error will occur.

The entry of calibration values outside the measuring range leads to a

calibration error.

1.

Bring the sensor into the measuring position.

2.

Switch to the measured value display with .

3.

Wait for a measurement, read the raw value (without dimen-

sion) and note it down. In the case of fluctuating values,

observe several measuring cycles and average the individual

raw values.

4.

Take a sample at the same time as the measurement and as

close to the UV-VIS sensor as possible.

5.

Determine the reference value and note it down as a value pair

together with the raw value.

6.

For a two-point calibration:

Determine the second value pair (step 4 to step 6).

7.

Sort the values pairs in ascending order.

8.

In the

Settings of sensors and diff. sensors

menu, enter the

value pair or pairs (see the following table).