No signal, Signal intensity is not constant, Peak wavelength has shifted – Ocean Optics ARCoptix ANIR User Manual
Page 29
4: Troubleshooting
ARC-00000-000-02-0310 21
No Signal
Possible Cause
Comment
Suggested Solution
USB cable is loose
Check cables for good contact
Light not on
Check light
Sample not appropriate
Use a different sample
Transient internal
spectrometer fault
Remove USB cable and re-insert
Defect in Spectrometer
Contact Ocean Optics
Signal Intensity is Not Constant
Possible Cause
Comment
Suggested Solution
Fiber optic cable has shifted
in position
Tape fiber optic cable to bench
Lamp is not stable
Defective lamp
Replace the lamp
Sample is not
homogeneous (cuvette)
Mix sample and filter if necessary to
remove particulate matter
Sample is not
homogeneous (integrating
sphere)
Move integrating sphere to a
different region of the sample
Sample is decomposing
Replace sample
Cuvette is scratched
Always place cuvette in
sample holder with same
plane facing the source
Replace cuvette
Defect in Spectrometer
Contact Ocean Optics
Peak Wavelength Has Shifted
Cause
Comment
Solution
The background spectrum
is not valid
Run a new background spectrum
Sample is not
homogeneous
Find another region of the sample
Defect in Spectrometer
Contact Ocean Optics