4: Troubleshooting

ARC-00000-000-02-0310 21

No Signal

Possible Cause

Comment

Suggested Solution

USB cable is loose

Check cables for good contact

Light not on

Check light

Sample not appropriate

Use a different sample

Transient internal
spectrometer fault

Remove USB cable and re-insert

Defect in Spectrometer

Contact Ocean Optics

Signal Intensity is Not Constant

Possible Cause

Comment

Suggested Solution

Fiber optic cable has shifted
in position

Tape fiber optic cable to bench

Lamp is not stable

Defective lamp

Replace the lamp

Sample is not
homogeneous (cuvette)

Mix sample and filter if necessary to
remove particulate matter

Sample is not
homogeneous (integrating
sphere)

Move integrating sphere to a
different region of the sample

Sample is decomposing

Replace sample

Cuvette is scratched

Always place cuvette in
sample holder with same
plane facing the source

Replace cuvette

Defect in Spectrometer

Contact Ocean Optics

Peak Wavelength Has Shifted

Cause

Comment

Solution

The background spectrum
is not valid

Run a new background spectrum

Sample is not
homogeneous

Find another region of the sample

Defect in Spectrometer

Contact Ocean Optics