4: Troubleshooting

20 ARC-000000-000-02-0310

As an example, if you change the supplier of a solvent, verify that the change has no effect on
the analysis.

 Run a standard (or a series of standards) on a periodic basis to verify that the system is

performing properly. When abnormal data is obtained, collect the spectrum of a standard and
compare the spectrum to one that was previously obtained.

 If an error message is presented, press the Continue button indicated on the Error Message

dialog box. If this does not remedy the problem, turn off the spectrometer, disconnect the
USB cable and then reinstall the cable.

The following SpectraSuite error message can be caused by the Windows firewall, which
blocks applications from communicating over the Internet. If this is the cause, follow the steps
given in the

Modifying Windows Firewall Settings

section. This error may also occur because

you have entered the wrong IP address or port number (see

Starting the SpectraSuite-Hub

Service

).

Spectrum is Noisy

Possible Cause

Comment

Suggested Solution

Lamp is weak

The lamp has a finite
lifetime

Replace the lamp

Prepare a more concentrated
sample

The absorbance is weak or
% T is very high (noisy
spectrum)

The sample is too dilute or
too thin

Use Spectrum Averaging to reduce
noise

The absorbance is too

strong

The sample is too
concentrated or thick

Dilute the sample or reduce its
thickness

Cable is loose

Check all cables for good contact

Fiber optic cable is
defective

Replace fiber optic cable

Transient internal
spectrometer fault

Remove USB cable and re-insert

Defect in Spectrometer

Contact Ocean Optics