2: Installing the HR2000+

294-00000-000-02-201012

7

Procedure

To configure hardware in USB Mode:

1. Specify S2000 in the Spectrometer Type drop-down menu.

2. Specify HR2000+ in the A/D Converter Type drop-down menu.

3. Specify the serial number of the HR2000+ under the USB Serial Number drop-down menu.

Note

The system pre-fills this drop-down menu with the serial numbers of all discovered
HR2000+ Spectrometers.

4. Click the OK button to accept the selected options.

5. The spectrometer should now be able to acquire data and respond to light. Exit and restart

OOIBase32 to save configuration data to disk.

Spectrometer Configuration Screen

The Spectrometer Configuration screen prompts you to configure specific channel-level spectrometer
information, if necessary.

Procedure

1. Select Spectrometer | Configure from the menu and set system parameters.

2. Select the Wavelength Calibration tab. OOIBase32 pre-fills the coefficients for the HR2000+

from information on a memory chip in the spectrometer.