Chapter 4: Experiments

22

000-20000-400-01-201108

Note

If you change the integration time or anything else in the experiment setup, you will need
to store new dark and reference readings.

2. Store the dark spectrum. Block the light by turning the mirror. The spectrum line will be flat

and close to the baseline. Click the dark spectrum bulb icon (

). The dark spectrum is now

stored.

3. Store the reference spectrum. Insert the reference cuvette into the sampling lamp. The reference

cuvette should contain only the solvent, usually water or an organic solvent. With a compound in

solution, the absorbance will reduce the peak. Then, click the Reference bulb icon (

).

The line can be displayed in 1

st

or 2

nd

order and forced through the origin. Values for the molar

extinction coefficient and R

2

are calculated.