Sievers 860 Laboratory System Specifications
1
Total Organic Carbon
Linear Range
0.3 ppb–1.5 ppm TOC
Precision
0.1 ppb or 1%, whichever is greater
Accuracy
± 5% at 500 ppb C of sucrose
Analysis Time
6 minutes
Sample Flow Rate (nominal)
Analysis mode: 0.25 mL/min
Fast Flush (between samples): 0.5 mL/min
Sample pH
<3
Sampling Method
Grab sampling
Operation
PC or standalone operation; one PC can control up to four analyzers
Analyzer Specifications
Power Requirements
100–240
±10% VAC, 60 W, 50/60 Hz
Fuses
T1.6 A, 250 VAC (SloBlo), size 5 x 20 mm (appliance inlet)
Ambient Temperature
10 °C to 40 °C
Normal Operating Environment
Intended for indoor use only
Maximum Relative Humidity
Up to 95%, non-condensing
Maximum Altitude
2
3,000 m (9,843 ft)
Sparge Gas (if used)
Inlet pressure: 30 psig to 50 psig
Sparge Gas Flow Rate
120 mL/min to 150 mL/min
Calibration Stability
Typically stable for 12 months
Outputs
One USB port to support printer or Flash drive, and one Ethernet port
Installation/Overvoltage Category
II
Safety Certifications
CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA C22.2
No. 61010-1
Pollution Degree
2
Display
Color LCD
Dimensions
H: 43.0 cm (16.9 in); W: 21.5 cm (8.5 in); D: 40.5 cm (15.9 in)
Weight
10.5 kg (23.0 lb)
* Trademark of General Electric Company; may be registered in one or more countries.
1
Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
2
Known to meet performance specifications
300 00240 Rev A
MC12-038
ISO
9001:2008
QUALITY
MANAGEMENT
SYSTEM
The Americas
GE Analytical Instruments
6060 Spine Road
Boulder, CO 80301-3687 USA
T +1 800 255 6964
T +1 303 444 2009
F +1 303 527 1797
[email protected]
Europe/Middle East/Africa
GE Analytical Instruments
Unit 3, Mercury Way
Urmston, Manchester
UK M41 7LY
T +44 (0) 161 864 6800
F +44 (0) 161 864 6829
[email protected]
Asia
[email protected]
[email protected]
[email protected]
For specific country contacts, visit
www.geinstruments.com/sales
www.geinstruments.com
© 2012
,
General Electric Company. All rights reserved.