beautypg.com

Additional information, For more information – GE P&W Pleated Filters - Memtrex NY Capsule Filters with Hydrophilic Nylon66 Membrane, Absolute User Manual

Page 2

background image

Page 2

FS1101EN

Table 2: Operational Limits

Description

Operational Limits

Max. Operational Pressure

80 psi (5.5 bar) at 70ºF (21ºC) in liquid
55 psi (3.8 bar) at 70ºF (21ºC) in gas

Max. Differential Pressure

60 psi (4.1 bar) at 70ºF (21ºC)

Max. Operating Temperature

110ºF (43ºC) at ≤30 psid (2.1 bar)
Operating Pressure

Table 3: Integrity Testing

Air diffusion through water saturated membrane is ≤1.8 cc/min per ft

2

of Effective Filtration Area

Micron Rating

Test Pressure

0.1 µm

40 psi (2.8 bar)

0.2 µm

30 psi (2.1 bar)

0.45 µm

20 psi (1.4 bar)

0.65 µm

10 psi (0.6 bar)

Additional Information

Memtrex NY Capsule filters may be sanitized with a
variety of commonly used chemical agents; how-
ever, acidic solutions should be avoided. The cap-
sules may be repeatedly autoclaved (250°F [121°C],
30 minutes) for up to 5 cycles.

GE certifies that the materials contained in its Mem-
trex NY Capsule filters meet U.S. FDA requirements
for food contact under the applicable regulations in
21 CFR. For further information, contact the GE
Technical Services Department. Memtrex NY filters
meet the test criteria for USP class VI-121°C Plas-
tics.

Figures 2, 3 and 4 show the test data results of flow
performance for the small, medium and large Mem-
trex NY Capsule Filters using a 10-inch length filter.

GE filter capsules are designed and manufactured
for resistance to a wide range of chemical solutions.
Conditions will vary with each application and users
should carefully verify chemical compatibility. Con-
sult the ordering information in Table 4 below or
contact your GE distributor for more information.

Figure 2: Memtrex NY small capsule flow rate perform-

ance in clean water.

2

Figure 3: Memtrex NY medium capsule flow rate per-

formance in clean water.

2

Figure 4: Memtrex NY large capsule flow rate perform-

ance in clean water.

2

2

Data based on 10 inch length filter

For more information

Email: [email protected]
• Tel: 800-444-8212 USA
• Canada Tel: 952-988-6665
• Web: www.geosmolabstore.com

Table 4: Ordering Information

Type

Absolute

Micron Rating

Capsule Size

Connections

3

CMNY

91 = 0.1 µm
92 = 0.2 µm

94 = 0.45 µm
96 = 0.65 µm

08 = small (0.8 ft

2

)

30 = medium (3.0 ft

2

)

59 = large (5.9 ft

2

)

L =

¼” – ½” (6.3 – 12.7 mm) hose barb

M =

3/8” (9.5 mm) hose barb

R =

¼” (6.3 mm) NPT mole

W =

½” (12.7 mm) NPT mole

Y=

1.5” (38.1 mm) sanitary flange

3

Choose an inlet and outlet connection