FS1101EN
Table 2: Operational Limits
Description
Operational Limits
Max. Operational Pressure
80 psi (5.5 bar) at 70ºF (21ºC) in liquid
55 psi (3.8 bar) at 70ºF (21ºC) in gas
Max. Differential Pressure
60 psi (4.1 bar) at 70ºF (21ºC)
Max. Operating Temperature
110ºF (43ºC) at ≤30 psid (2.1 bar)
Operating Pressure
Table 3: Integrity Testing
Air diffusion through water saturated membrane is ≤1.8 cc/min per ft
2
of Effective Filtration Area
Micron Rating
Test Pressure
0.1 µm
40 psi (2.8 bar)
0.2 µm
30 psi (2.1 bar)
0.45 µm
20 psi (1.4 bar)
0.65 µm
10 psi (0.6 bar)
Additional Information
Memtrex NY Capsule filters may be sanitized with a
variety of commonly used chemical agents; how-
ever, acidic solutions should be avoided. The cap-
sules may be repeatedly autoclaved (250°F [121°C],
30 minutes) for up to 5 cycles.
GE certifies that the materials contained in its Mem-
trex NY Capsule filters meet U.S. FDA requirements
for food contact under the applicable regulations in
21 CFR. For further information, contact the GE
Technical Services Department. Memtrex NY filters
meet the test criteria for USP class VI-121°C Plas-
tics.
Figures 2, 3 and 4 show the test data results of flow
performance for the small, medium and large Mem-
trex NY Capsule Filters using a 10-inch length filter.
GE filter capsules are designed and manufactured
for resistance to a wide range of chemical solutions.
Conditions will vary with each application and users
should carefully verify chemical compatibility. Con-
sult the ordering information in Table 4 below or
contact your GE distributor for more information.
Figure 2: Memtrex NY small capsule flow rate perform-
ance in clean water.
2
Figure 3: Memtrex NY medium capsule flow rate per-
formance in clean water.
2
Figure 4: Memtrex NY large capsule flow rate perform-
ance in clean water.
2
2
Data based on 10 inch length filter
For more information
• Email: [email protected]
• Tel: 800-444-8212 USA
• Canada Tel: 952-988-6665
• Web:
Table 4: Ordering Information
Type
Absolute
Micron Rating
Capsule Size
Connections
3
CMNY
91 = 0.1 µm
92 = 0.2 µm
94 = 0.45 µm
96 = 0.65 µm
08 = small (0.8 ft
2
)
30 = medium (3.0 ft
2
)
59 = large (5.9 ft
2
)
L =
¼” – ½” (6.3 – 12.7 mm) hose barb
M =
3/8” (9.5 mm) hose barb
R =
¼” (6.3 mm) NPT mole
W =
½” (12.7 mm) NPT mole
Y=
1.5” (38.1 mm) sanitary flange
3
Choose an inlet and outlet connection