Additional information, For more information – GE P&W Pleated Filters - Memtrex MP Capsule Filters with Polyethersulfone Membrane, Absolute User Manual
Page 2

Table 2: Operational Limits
Page 2
FS1104EN
Description
Operational Limits
Max. Operational Pressure
80 psi (5.5 bar) at 70ºF (21ºC) in liquid
55 psi (3.8 bar) at 70ºF (21ºC) in gas
Max. Differential Pressure
60 psi (4.1 bar) at 70ºF (21ºC)
Max. Operating Temperature
110ºF (43ºC) at ≤30 psid (2.1 bar)
Operating Pressure
Table 3: Integrity Testing
Air diffusion through water saturated membrane
Micron Rating
Test Pressure
Specification
0.1 μm
50 psi (3.4 bar)
≤5.6 cc/min per ft of EFA
2
0.2 μm
30 psi (2.1 bar)
≤2.4 cc/min per ft of EFA
2
0.45 μm
20 psi (1.4 bar)
≤2.0 cc/min per ft of EFA
2
0.65 μm
13 psi (0.9 bar)
≤1.5 cc/min per ft of EFA
2
Additional Information
Memtrex MP Capsule filters may be sanitized with a
variety of commonly used chemical agents. The
capsules may be repeatedly autoclaved (250°F
[121°C], 30 minutes) for up to 5 cycles.
GE certifies that the materials contained in its
Memtrex MP Capsule filters meet U.S. FDA require-
ments for food contact under the applicable regula-
tions in 21 CFR. For further information, contact the
GE Technical Services Department. Memtrex MP
filters meet the test criteria for USP class VI-121°C
Plastics.
Figures 2, 3 and 4 show the test data results of flow
performance for the small, medium and large Mem-
trex MP Capsule Filters using a 10-inch length filter.
GE filter capsules are designed and manufactured
for resistance to a wide range of chemical solutions.
Conditions will vary with each application and users
should carefully verify chemical compatibility. Con-
sult Table 4 below for ordering information or con-
tact your GE distributor for more information.
Figure 2: Memtrex MP small capsule flow rate per-
formance in clean water.
2
Figure 3: Memtrex MP medium capsule flow rate per-
formance in clean water.
2
Figure 4: Memtrex MP large capsule flow rate perform-
ance in clean water.
2
2
Data based on 10 inch length filter
For more information
• Email: labstor[email protected]
• Tel: 800-444-8212 USA
• Canada Tel: 952-988-6665
• Web: www.geosmolabstore.com
Table 4: Ordering Information
Type
Absolute
Micron Rating
Capsule Size
Connections
3
CMMP
91 = 0.1 μm
92 = 0.2 μm
94 = 0.45 μm
96 = 0.65 μm
08 = small (0.8 ft
2
)
30 = medium (3.0 ft )
2
59 = large (5.9 ft )
2
L =
¼” – ½” (6.3 – 12.7 mm) hose barb
M =
3/8” (9.5 mm) hose barb
R =
¼” (6.3 mm) NPT mole
W =
½” (12.7 mm) NPT mole
Y=
1.5” (38.1 mm) sanitary flange
Choose an inlet and outlet connection
3