Table 2: Operational Limits

Page 2

FS1104EN

Description

Operational Limits

Max. Operational Pressure

80 psi (5.5 bar) at 70ºF (21ºC) in liquid
55 psi (3.8 bar) at 70ºF (21ºC) in gas

Max. Differential Pressure

60 psi (4.1 bar) at 70ºF (21ºC)

Max. Operating Temperature

110ºF (43ºC) at ≤30 psid (2.1 bar)
Operating Pressure

Table 3: Integrity Testing

Air diffusion through water saturated membrane

Micron Rating

Test Pressure

Specification

0.1 μm

50 psi (3.4 bar)

≤5.6 cc/min per ft of EFA

2

0.2 μm

30 psi (2.1 bar)

≤2.4 cc/min per ft of EFA

2

0.45 μm

20 psi (1.4 bar)

≤2.0 cc/min per ft of EFA

2

0.65 μm

13 psi (0.9 bar)

≤1.5 cc/min per ft of EFA

2

Additional Information

Memtrex MP Capsule filters may be sanitized with a
variety of commonly used chemical agents. The
capsules may be repeatedly autoclaved (250°F
[121°C], 30 minutes) for up to 5 cycles.

GE certifies that the materials contained in its
Memtrex MP Capsule filters meet U.S. FDA require-
ments for food contact under the applicable regula-
tions in 21 CFR. For further information, contact the
GE Technical Services Department. Memtrex MP
filters meet the test criteria for USP class VI-121°C
Plastics.

Figures 2, 3 and 4 show the test data results of flow
performance for the small, medium and large Mem-
trex MP Capsule Filters using a 10-inch length filter.

GE filter capsules are designed and manufactured
for resistance to a wide range of chemical solutions.
Conditions will vary with each application and users
should carefully verify chemical compatibility. Con-
sult Table 4 below for ordering information or con-
tact your GE distributor for more information.

Figure 2: Memtrex MP small capsule flow rate per-

formance in clean water.

2

Figure 3: Memtrex MP medium capsule flow rate per-

formance in clean water.

2

Figure 4: Memtrex MP large capsule flow rate perform-

ance in clean water.

2

2

Data based on 10 inch length filter

For more information

• Email: labstor[email protected]
• Tel: 800-444-8212 USA
• Canada Tel: 952-988-6665
• Web: www.geosmolabstore.com

Table 4: Ordering Information

Type

Absolute

Micron Rating

Capsule Size

Connections

3

CMMP

91 = 0.1 μm
92 = 0.2 μm

94 = 0.45 μm
96 = 0.65 μm

08 = small (0.8 ft

2

)

30 = medium (3.0 ft )

2

59 = large (5.9 ft )

2

L =

¼” – ½” (6.3 – 12.7 mm) hose barb

M =

3/8” (9.5 mm) hose barb

R =

¼” (6.3 mm) NPT mole

W =

½” (12.7 mm) NPT mole

Y=

1.5” (38.1 mm) sanitary flange

Choose an inlet and outlet connection