5 Connecting to voltage supply

5.1 Preparing the connection

Generally not the following safety instructions:

l

Connect only in the complete absence of line voltage

l

If overvoltage surges are expected, overvoltage arresters

should be installed according to Foundation Fieldbus

specification.

In hazardous areas you should take note of the appropriate

regulations, conformity and type approval certificates of the

sensors and power supply units.

VEGASON 66 requires a supply voltage of 9 … 32 V DC.

Power and the digital bus signal are carried on the same two-

wire connection cable. Power is supplied via the H1 power

supply.
VEGASON 66 is connected with screened cable according to

the Fieldbus specification. An outer cable diameter of 5 … 9 mm

ensures the seal effect of the cable gland.
Please make sure that your entire installation is carried out

according to the Fieldbus specification. In particular, make

sure that the termination of the bus is done with appropriate

terminating resistors.
On VEGASON 66 with cable gland ½ NPT and plastic housing,

a metal ½" threaded insert is moulded in the plastic housing.

Caution:

No grease should be used when screwing the NPT cable

gland or steel tube into the threaded insert. Standard grease

can contain additives affecting the connection between

threaded insert and housing. This will influence the stability of

the connection and the tightness of the housing.

In systems with potential equalisation, connect the cable

screen directly to ground potential at the power supply unit, in

the connection box and at the sensor. The screen in the sensor

must be connected directly to the internal ground terminal. The

ground terminal outside on the housing must be connected to

the potential equalisation (low impedance).

Note safety instructions

Take note of safety

instructions for Ex

applications

Select power supply

Selecting the connection ca-

ble

Cable gland ½ NPT

Cable screening and ground-

ing