The reference plane for the measuring range (version A) is the

lower edge of the flange. The reference plane for versions B, C

and D is the lower edge of the transducer.
Make sure that for all versions a min. distance - the so called

dead zone - below the reference plane is maintained in which a

measurement is not possible. The exact value of the dead

zone is stated in the "Technical data" in the "Supplement".

Information:

If the medium reaches the transducer, buildup can form on it

and cause faulty measurements later on.

1

3

2

100%

0%

4

Fig. 3: Version A - Measuring range (operating range) and max. measuring

distance

1

full (dead zone)

2

empty (max. measuring distance)

3

Measuring range

4

Reference plane

Measuring range

12

VEGASON 66 - Foundation Fieldbus

Mounting

28795

-EN