VEGA VEGABAR 61 4 … 20 mA_HART User Manual
Page 64
1000
750
500
250
12
18
16
14
20
22
24
26
28
30
32
34
36
Ω
V
4
1
2
3
Fig. 30: Voltage diagram
1
HART load
2
Voltage limit EEx-ia instrument
3
Voltage limit non-Ex/Exd instrument
4
Operating voltage
Electrical protective measures
P
rotection class
-
H
ousing, standard
IP
66/IP 67
20)
-
A
luminium and stainless housing (op-
tionally available)
IP
68 (1
bar)
21)
-
T
ransmitter in IP 68 version
IP
68
-
E
xternal housing
IP
65
O
vervoltage category
III
P
rotection class
II
Functional safety (SIL)
T
he functional safety is already activated Ex factory for instruments with SIL qualification. For
instruments Ex factory without SIL qualification, the functional safety must be activated by the user
for applications according to SIL via the indicating and adjustment module or via PACTware.
F
unctional safety according to IEC 61508-4
-
S
ingle channel architecture (1oo1D)
up to SIL2
-
double channel diversitary redundant
architecture (1oo2D)
up to SIL3
20)
I
nstruments with gauge pressure measuring ranges cannot detect the am-
bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.
21)
O
nly with instruments with absolute pressure ranges.
64
VEGABAR
61 • 4 … 20
mA/HART
10 S
upplement
27519
-EN
-081209