beautypg.com

VEGA VEGABAR 61 4 … 20 mA_HART User Manual

Page 64

background image

1000

750

500

250

12

18

16

14

20

22

24

26

28

30

32

34

36

V

4

1

2

3

Fig. 30: Voltage diagram
1

HART load

2

Voltage limit EEx-ia instrument

3

Voltage limit non-Ex/Exd instrument

4

Operating voltage

Electrical protective measures

P

rotection class

-

H

ousing, standard

IP

66/IP 67

20)

-

A

luminium and stainless housing (op-

tionally available)

IP

68 (1

bar)

21)

-

T

ransmitter in IP 68 version

IP

68

-

E

xternal housing

IP

65

O

vervoltage category

III

P

rotection class

II

Functional safety (SIL)

T

he functional safety is already activated Ex factory for instruments with SIL qualification. For

instruments Ex factory without SIL qualification, the functional safety must be activated by the user
for applications according to SIL via the indicating and adjustment module or via PACTware.

F

unctional safety according to IEC 61508-4

-

S

ingle channel architecture (1oo1D)

up to SIL2

-

double channel diversitary redundant
architecture (1oo2D)

up to SIL3

20)

I

nstruments with gauge pressure measuring ranges cannot detect the am-

bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.

21)

O

nly with instruments with absolute pressure ranges.

64

VEGABAR

61 • 4 … 20

mA/HART

10 S

upplement

27519

-EN

-081209