VEGA VEGABAR 61 4 … 20 mA_HART User Manual
Page 57
A
djustment range of the min./max. adjustment relating to the nominal measuring range:
-
percentage value
-10 … 110 %
-
pressure value
-20 … 120 %
A
djustment range of the zero/span adjustment relating to the nominal measuring range:
-
zero
-20 … +95 %
-
S
pan
-120 … +120 %
9)
-
D
ifference between zero and span
max. 120 % of the nominal range
A
djustment range with measuring ranges from 100 bar, relating to the nominal measuring range:
-
zero/Min.
-5 … +95 %
-
span/Max.
-5 … +105 %
R
ecommended max. turn down
10 : 1 (
no limitation)
Nominal measuring ranges and overload resistance
Nominal range
Overload, max. pressure Overload, min. pressure
G
auge pressure
0 … 0.4
bar/0 … 40 kPa
30
bar/3000 kPa
-0.8
bar/-80 kPa
0 … 1
bar/0 … 100 kPa
35
bar/3500 kPa
-1
bar/-100 kPa
0 … 2.5
bar/0 … 250 kPa
50
bar/5000 kPa
-1
bar/-100 kPa
0 … 100
bar/0 … 10 MPa
200
bar/20000 kPa
-1
bar/-100 kPa
0 … 250
bar/0 … 25 MPa
500
bar/50 MPa
-1
bar/-100 kPa
0 … 400
bar/0 … 40 MPa
800
bar/80 MPa
-1
bar/-100 kPa
0 … 5
bar/0 … 500 kPa
65
bar/6500 kPa
-1
bar/-100 kPa
0 … 10
bar/0 … 1000 kPa
90
bar/9000 kPa
-1
bar/-100 kPa
0 … 25
bar/0 … 2500 kPa
130
bar/13000 kPa
-1
bar/-100 kPa
0 … 60
bar/0 … 6000 kPa
200
bar/20000 kPa
-1
bar/-100 kPa
-1 … 0
bar/-100 … 0 kPa
35
bar/3500 kPa
-1
bar/-100 kPa
-1 … 1.5
bar/-100 … 150 kPa
50
bar/5000 kPa
-1
bar/-100 kPa
-1 … 5
bar/-100 … 500 kPa
65
bar/6500 kPa
-1
bar/-100 kPa
-1 … 10
bar/-100 … 1000 kPa
90
bar/9000 kPa
-1
bar/-100 kPa
-1 … 25
bar/-100 … 2500 kPa
130
bar/13000 kPa
-1
bar/-100 kPa
-1 … 60
bar/-100 … 6000 kPa
200
bar/20000 kPa
-1
bar/-100 kPa
-0.2 … 0.2
bar/-20 … 20 kPa
30
bar/3000 kPa
-1
bar/-100 kPa
-0.5 … 0.5
bar/-50 … 50 kPa
35
bar/3500 kPa
-1
bar/-100 kPa
0 … 2.5
bar/0 … 250 kPa
50
bar/5000 kPa
-1
bar/-100 kPa
A
bsolute pressure
0 … 1
bar/0 … 100 kPa
35
bar/3500 kPa
0
bar abs.
0 … 2.5
bar/0 … 250 kPa
50
bar/5000 kPa
0
bar abs.
0 … 5
bar/0 … 500 kPa
65
bar/6500 kPa
0
bar abs.
0 … 10
bar/0 … 1000 kPa
90
bar/9000 kPa
0
bar abs.
9)
V
alues less than -1 bar cannot be set.
VEGABAR
61 • 4 … 20
mA/HART
57
10 S
upplement
27519
-
EN
-081209