beautypg.com

VEGA VEGABAR 61 4 … 20 mA_HART User Manual

Page 59

background image

Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)

A

pplies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well as to analogue

current output 4 … 20 mA. Specifications refer to the set span. Turn down (TD) = nominal
measuring range/set span.

L

ong-term drift of the zero signal

<

(0.1 %

x TD)/year

Ambient conditions

A

mbient, storage and transport temperature

-

S

tandard version

-40 … +80 °C (-40 … +176 °F)

-

V

ersion for oxygen applications

11)

-40 … +60 °C (-40 … +140 °F)

-

V

ersions IP 66/IP 68, 1 bar and IP 68, PE

connection cable

-20 … +60 °C (-4 … +140 °F)

-

V

ersions IP 66/IP 68, 1 bar and IP 68,

PUR

connection cable

-20 … +80 °C (-4 … +176 °F)

Additional temperature influence by isolating diaphragm

T

he specifications refer to diaphragm material 316L as well as isolating liquid silicone oil. They are

only used for estimation. The actual values depend on the diameter, material and strength of the
diaphragm as well as the isolating liquid. They are available on request.

T

emperature coefficient of the isolating diaphragm in mbar/10 K with

-

F

lange DN 25 PN 40, Form C, DIN 2501

4.8

-

F

lange DN 40 PN 40, Form C, DIN 2501

1

-

F

lange DN 50 PN 40, Form C, DIN 2501

2

-

F

lange DN 50 PN 40 with extension

50

mm

1.9

-

F

lange DN 50 PN 40 with extension

100

mm

1.9

-

F

lange DN 50 PN 40 with extension

200

mm

2.1

-

F

lange DN 80 PN 40, Form C, DIN 2501

0.4

-

F

lange 1" 150 lbs RF ANSI B16.5; with

extension 2", 3"

1.8

-

F

lange 2" 150 lbs RF ANSI B16.5; with

extension 2"

2

T

emperature coefficient of a cooling ele-

ment, depending on the diaphragm-ø

0.1 … 1.5

T

emperature coefficient of a 1 m long

capillary line, depending on the diaphragm-ø

0.1 … 15

11)

U

p to 60 °C (140 °F).

VEGABAR

61 • 4 … 20

mA/HART

59

10 S

upplement

27519

-

EN

-081209