beautypg.com

VEGA VEGABAR 61 4 … 20 mA_HART User Manual

Page 38

background image

T

he position correction compensates the influence of the mounting

position or static pressure on the measurement. It does not influence
the adjustment values.

I

n the menu items "zero" and "span" you determine the span of the

sensor, the span corresponds to the end value.

Information:
T

he steps 1, 3 and 4 are not necessary for instruments which are

already preset according to customer specifications!

Y

ou can find the data on the type label on the instrument or in the

menu items of the zero/span adjustment.

T

he indicating and adjustment module enables the adjustmetn without

lling or pressure. Thanks to this, you can carry out your settings

already in the factory without the instrument having to be installed.

T

he actual measured value is displayed in addition to the menu items

for zero/span adjustment.

VEGABAR

61

is preset to application "Level measurement". Proceed

as follows when switching over to application "Process pressure
measurement":

1

P

ush the [OK] button in the measured value display, the menu

overview is displayed.

2

S

elect the menu "Service" with [->] and confirm with [OK].

B

asic adjustment

D

isplay

D

iagnostics

S

ervice

I

nfo

3

S

elect the menu item "Application" with [->] and edit with [OK].

Warning:
N

ote the warning: "Output can change".

4

S

elect with [->] "OK" and confirm with [OK].

5

S

elect "Process pressure" from the list and confirm with [OK].

I

n this menu item you select the adjustment unit as well as the unit for

the temperature indication in the display.

T

o select the adjustment unit (in the example switching over from bar

to mbar) you have to proceed as follows:

5)

1

P

ush the [OK] button in the measured value display, the menu

overview is displayed.

5)

S

election options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH

2

O,

mmH

2

O.

Select application
"P

rocess pressure

measurement"

Select unit

38

VEGABAR

61 • 4 … 20

mA/HART

6 S

et up with the indicating and adjustment module PLICSCOM

27519

-EN

-081209