2 approvals, 3 features, Ex approval – VEGA VEGAMET 513 User Manual
Page 5: Whg approval
VEGAMET 513
5
Product description
1.2 Approvals
VEGAMET are available with the following
approvals:
- explosion protection: appropriate operating
instrument with intrinsically safe circuit
(VEGAMET 513 Ex)
- overfill protection acc. to WHG
For such applications, please note the appro-
priate official documents (general type ap-
proval, test and conformity certificates). They
come with the respective instrument
Ex approval
For hazardous areas, certification acc. to
CENELEC defined in the conformity certificate
PTB no. Ex-95.D.2145 X for VEGAMET 513
Ex.
WHG approval
Signal conditioning instrument as part of an
overfill protection system acc. to WHG (test
certificate number see table).
1.3 Features
- microcomputer-controlled signal condition-
ing instrument for continuous measurement
- adjustment module with LC display and 6
keys
- adjustable integration time
- two fixed and three user-programmable
linearisation curves
- fault monitoring
- 1 sensor input (capacitive electrode or
pressure transmitter)
- fault signal and failure diagnosis via dis-
play
- 1 current output 0/4 … 20 mA (floating)
- 1 voltage output 0/2 … 10 V
- 1 DISBUS output for digital wiring and for
connection of VEGADIS 174
- 1 fail-safe relay
- adjustment also via PC with adjustment
software VVO; connection via the interface
adapter of the VEGACONNECT series
Instrument
Oscillator
VEGAMET
Test certificate
Auxiliary level switch
no.
VEGASEL
Capacitive electrodes
E17 EX
513 EX
Z-65.13-123
543 … 547
EL 11 EX0, EL 21 EX0,
E18 EX
643
EL 24 EX0, EL 29 EX,
EL 31 EX0, EL 42 EX0
Pressure transmitters
E23,
513
Z-65.11-130
543 … 547
D80, D81, D84, D86,
E23 Ex
513 EX
643
D84 EX, D86 EX
Pressure transmitters
E22,
513
Z-65.11-130
543 … 547
D77, D85, D87,
E22 Ex
513 Ex
643
D77 EX, D85 EX, D87 EX