6 settings in the "reduced menu, 1 configuration of measurement loop – VEGA VEGAMET 513 User Manual
Page 19
VEGAMET 513
19
Settings in the "reduced menu"
Name measurement loop
You can modify the meas-
urement loop name "TAG-
No. 1", if you want a more
useful designation.
Two lines with six figures
each are available. To do
this, move to the menu item
"TAG name …".
Push [+] or [–] once. The first char-
acter of the measurement loop name,
in this example the letter T, flashes
(editing mode). Continue pushing the
[+] or [–] key, by doing this you scroll
through the alphabet, through the row of
numbers and through a list of characters.
When you have reached the re-
quested character, push [–>]. The
next character is activated and
begins flashing. Scroll again with [+]
and [–] to the requested character. With [–>]
you again move to the next character.
When you have finished the last
position of the name and you have
"written" the name of the measure-
ment loop in this way, you have to
push [OK] to save the setting. If you quit the
menu item without saving, the name of the
measurement loop is again "TAG-No. 1".
In the display of measured value and
all other menu items in which the
measurement loop name is shown,
you will now see the new designation.
TAG-
No. 1
%
xx,x
Param.
TAG-
No. 1
Confi-
gura-
tion
TAG
name
TAG -
No. 1
TAG
name
T
AG -
No. 1
TAG
name
T
A
G -
No. 1
TAG
name
Ves.4
Hall
3
Ves.4
Hall3
%
27,8
6 Settings in the "reduced menu"
6.1 Configuration of measurement loop
Choose sensor type
By default, VEGAMET 513 is configured for
connection to a hydrostatic sensor. If, for
example, you use a capacitive electrode, you
have to proceed as follows:
Move to the menu item "Sensor type
Hydrostatic".
Push [+] or [–] once. The word "Hydrostatic"
flashes. You are now in the editing mode.
Continue pushing [+] or [–] until the display
shows "Sensor type capacitive". You are still
in the editing mode.
Now push the [OK] key. By doing this, you
save the settings and quit the editing mode.
Meas. value
indication
TAG-
No. 1
%
xx,x
Param.
TAG-
No. 1
Confi-
gura-
tion
Add’l
func-
tions
Pass-
word
off
Lang
uage
Eng-
lish
TAG 1
to de-
fault
Sensor
type
Hydro-
static
Reset
TAG 1
Delete
TAG
1?
It is only possible to
modify the sensor type,
when a "Reset TAG 1" was
carried out before.
Reset
OK ?
Reset
Now !
OK ?
capa-
citive
VEGADIF