4 adjustment – VEGA VEGAMET 513 User Manual
Page 29
VEGAMET 513
29
7.4 Adjustment
The
adjustment with medium
corresponds
to the procedure in the reduced menu (see
appropriate chapter).
- if the medium is changed later on (medium
with another density value or DK value)
and no new adjustment can be made
or
- if the MET indication and the DIS outputs
relate to height in meters,
Please note the following:
In applications with hydrostatic pressure
transmitters, the adjusted density value (see
menu Parameter - Conditioning - Scaling -
Density) must correspond to the value of the
medium. In applications with capacitive elec-
trodes, the adjusted DK value must corre-
spond to the value of the medium.
Settings in the "extended menu"
Param.
TAG-
No. 1
Adjust-
ment
with
medium
without
medium
Min-
adjust
at %
0,0
Max-
adjust
at %
100,0
Adjust-
ment
mA
Offset
cor-
rection
0%
at
mA
0,000
100%
at
mA
1,000
Sensor
unpres-
sur'd ?
OK ?
Offset
corr.
Now!
OK ?
TAG -
No. 1
%
xx,x
For the
adjustment without medium,
two
anticipated values for level, gauge, etc.
corresponding to 0 % and 100 % must be
entered.
The menu item "Offset correction" is opened
when hydrostatic pressure transmitters are
connected. The offset correction should be
carried out after installation of the sensor as
the sensor values can differ slightly due to
different installation positions (vertical or
horizontal installations). With the offset cor-
rection, the meas. data of the unpressurised
sensor (vessel empty) are detected as cor-
rection values and taken into account in all
subsequent measurements.