2 audit trail seal – Ohaus VX SERIES FLOOR SCALES Indicator T32XW Manual en User Manual
Page 35

EN-33
3000 Series Indicators
Fgure 6-1. Wre Seal Fgure 6-2. Paper Seal
SEAL
Seal
.3.2 Audit Trail Seal
For jursdctons that use the audt tral sealng method, the local weghts and measures offcal or authorzed servce agent must
record the current confguraton and calbraton event counter values at the tme of sealng. These values wll be compared to
values found durng a future nspecton.
NOTE: A change to an event counter value s equvalent to breakng a physcal seal.
The audt tral uses two event counters to record changes to confguraton and calbraton settngs.
The confguraton event counter (CFG) wll ndex by 1 when extng the menu f one or more of the followng settngs
are changed Legal for Trade, Calbraton Unt, Capacty, Graduaton, Power On Unt, Zero Range, Auto Zero Trackng,
Expanded Mode, Count Mode, Klogram Unt, Pound Unt, Gram Unt, Ounce Unt, Pound Ounce Unt, Stable Only. Note
that the counter only ndexes once, even f several settngs are changed. The confguraton event counter values range
from CFG000 to CFG999. When the value reaches CFG999, the count starts over at CFG000.
The calbraton event counter (CAL) wll ndex by 1 when extng the menu f a Span Calbraton, Lnearty Calbraton
or GEO settng change s made. Note that the counter only ndexes once, even f several settngs are changed. The
calbraton event counter values range from CAL000 to CAL999. When the value reaches CAL999, the count starts over
at CAL000.
The event counters can be vewed by pressng and holdng the MENU button.
Whle the button s held, the dsplay wll show MENU followed by Audt.
Release the button when Audt s dsplayed to vew the audt tral nformaton.
•
•