Legal for trade – Ohaus VX SERIES FLOOR SCALES Indicator T32XW Manual en User Manual
Page 34

EN-32
3000 Series Indicators
. LEGAL FOR TRADE
When the ndcator s used n trade or a legally controlled applcaton t must be set up, verfed and sealed n accordance wth
local weghts and measures regulatons. It s the responsblty of the purchaser to ensure that all pertnent legal requrements are
met.
. Settings
Before verfcaton and sealng, perform the followng steps:
1. Verfy that the menu settngs meet the local weghts and measures regulatons.
2. Perform a calbraton.
3. Set Legal for Trade to ON n the Setup menu.
4. Ext the menu.
5. Dsconnect power from the ndcator and open the housng as explaned n Secton 2.3.1.
6. Set the poston of the securty swtch to ON as shown n Secton 1.2, Fgure 1-2, Item 8.
7. Close the housng.
8. Reconnect power and turn the ndcator on.
NOTE: For nstallatons that employ the audt tral sealng method, steps 5 to 8 are not requred. However, the securty swtch
may be set to ON to safeguard aganst unntentonal changes to confguraton and calbraton settngs.
NOTE: When Legal for Trade s set to ON and the securty swtch s set to ON, the followng menu settngs cannot be changed:
Span Calbraton, Lnearty Calbraton, GEO, LFT, Calbraton Unt, Capacty, Graduaton, Power On Unt, Zero Range, Auto Zero
Trackng, Expanded Mode, Count Mode, Klogram Unt, Pound Unt, Gram Unt, Ounce Unt, Pound Ounce Unt, Stable Only. To
enable edtng of these menu settngs, return the securty swtch to the off poston and set LFT menu tem to off.
.2 Verification
The local weghts and measures offcal or authorzed servce agent must perform the verfcaton procedure. Please contact your
local weghts and measures offce for further detals.
.3 Sealing
.3. Physical Seals
For jursdctons that use the physcal sealng method, the local weghts and measures offcal or authorzed servce agent must
apply a securty seal to prevent tamperng wth the settngs. Refer to the llustratons below for sealng methods.