3000 Series Indicators
3.3. Span Calibration
Span Calbraton uses two ponts to adjust the scale. The frst pont s the zero value where there s no weght
on the scale. The second pont s the Span value where a calbraton mass s placed on the scale.
When SPAN s dsplayed, press the
Yes button to access the Span Calbraton menu tem.
The dsplay flashes 0.
Wth no weght on the scale, press the
Yes button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
The dsplay flashes the span calbraton pont. Place the specfed weght on the scale and press the
Yes
button.
To choose a dfferent span pont, repeatedly press the
No button to ncrement the selectons or press the Back
button to decrement the selectons. Refer to Table 3-3 for avalable span ponts. When the desred value s
dsplayed, place the specfed weght on the scale and press the
Yes button.
The dsplay shows --C-- whle the span pont s establshed.
If span calbraton was successful, the scale exts to the actve weghng mode and dsplays the actual weght
value.
3.3.2 Linearity Calibration
Lnearty calbraton uses 3 calbraton ponts. The frst calbraton pont s establshed wth no weght on the
scale. The second calbraton pont s establshed at approxmately half capacty. The thrd calbraton pont
s establshed at capacty. The Lnearty calbraton ponts are fxed and cannot be altered by the user durng
the calbraton procedure. Refer to Table 3-3 for the lnearty ponts.
When LINEAr s dsplayed, press the
Yes button to access the Lnearty Calbraton menu tem.
The dsplay flashes 0. Wth no weght on the scale, press the
Yes button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
The dsplay flashes the md calbraton pont.
Place the specfed weght on the scale and press the
Yes button.
The dsplay shows --C-- whle the md pont s establshed.
The dsplay flashes the full calbraton pont.
Place the specfed weght on the scale and press the
Yes button.
The dsplay shows --C-- whle the full pont s establshed.
If lnearty calbraton was successful, the scale exts to the actve weghng mode and dsplays the actual
weght value.