3000 Series Indicators
3.3. Span Calibration
Span	Calbraton	uses	two	ponts	to	adjust	the	scale.		The	frst	pont	s	the	zero	value	where	there	s	no	weght	
on	the	scale.		The	second	pont	s	the	Span	value	where	a	calbraton	mass	s	placed	on	the	scale.
When SPAN s dsplayed, press the
Yes button to access the Span Calbraton menu tem.
The dsplay flashes 0.
Wth no weght on the scale, press the
Yes button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
The dsplay flashes the span calbraton pont. Place the specfed weght on the scale and press the
Yes
button.
To choose a dfferent span pont, repeatedly press the
No button to ncrement the selectons or press the Back
button	to	decrement	the	selectons.		Refer	to	Table	3-3	for	avalable	span	ponts.		When	the	desred	value	s	
dsplayed,	place	the	specfed	weght	on	the	scale	and	press	the	
Yes button.
The dsplay shows --C-- whle the span pont s establshed.
If	span	calbraton	was	successful,	the	scale	exts	to	the	actve	weghng	mode	and	dsplays	the	actual	weght	
value.		
3.3.2 Linearity Calibration
Lnearty	calbraton	uses	3	calbraton	ponts.		The	frst	calbraton	pont	s	establshed	wth	no	weght	on	the	
scale.		The	second	calbraton	pont	s	establshed	at	approxmately	half	capacty.		The	thrd	calbraton	pont	
s	establshed	at	capacty.		The	Lnearty	calbraton	ponts	are	fxed	and	cannot	be	altered	by	the	user	durng	
the	calbraton	procedure.		Refer	to	Table	3-3	for	the	lnearty	ponts.
When LINEAr s dsplayed, press the
Yes button to access the Lnearty Calbraton menu tem.
The dsplay flashes 0. Wth no weght on the scale, press the
Yes button to establsh the zero pont.
The dsplay shows --C-- whle the zero pont s establshed.
The dsplay flashes the md calbraton pont.
Place the specfed weght on the scale and press the
Yes button.
The dsplay shows --C-- whle the md pont s establshed.
The dsplay flashes the full calbraton pont.
Place the specfed weght on the scale and press the
Yes button.
The dsplay shows --C-- whle the full pont s establshed.
If	lnearty	calbraton	was	successful,	the	scale	exts	to	the	actve	weghng	mode	and	dsplays	the	actual	
weght	value.