Ohaus VX SERIES FLOOR SCALES Indicator T32XW Manual en User Manual

Page 20

background image

EN-

3000 Series Indicators

Reset

No, Yes

Legal for Trade

Off, On

Cal Unt

kg, lb

Capacty

…20000

Graduaton

0.00…20

Power On Unt g, kg, lb, oz, lb:oz,

Auto

Zero Range 0%,

2%, 100%

End Setup Ext SETUP menu

3. Setup Menu

When the Indcator s used for the frst tme, enter ths
menu to set the Capacty and Graduaton.

3.. Reset

Reset the Setup menu to the factory defaults.
No = not reset.
Yes = reset.

NOTE: If the Legal for Trade menu tem s set to ON, the Capacty, Graduaton, Zero Range and Legal For
Trade settngs are not reset.

3.. Capacity

Set the scale capacty from 5 to 20000. Refer to the Setup Table 3.3 for avalable settngs.

3..2 Legal for Trade

Set the legal for trade status.
OFF = off
ON = on
Turnng on the “LFT” menu settng has the followng effects:
• Zero-range s set and locked on “2”.
• Auto Zero Trackng s set and locked on 0.5d
• The lb:oz unt s not avalable as a power-on settng.

3..3 Calibration Unit

Set the unt durng calbraton.
CAL UN kg = Calbrate usng kg weghts
CAL UN lb = Calbrate usng pound weghts