LINEARITY CALIBRATION
1. With the balance OFF, press Zero On until
MENU
is displayed.
2. Repeately press Mode Off until
Lin
is displayed, then press Zero On.
-C-
is momentarily displayed followed by the value of the mass which must
be placed on the pan.
Do not disturb the balance while
-C-
is displayed.
3. Place the displayed value of the first mass on the pan and momentarily
press Zero On.
-C-
is displayed, then the value of the second mass to be
placed on the pan is displayed.
4. Place the displayed value of the second mass on the pan and momentarily
press Zero On.
5. When the weight on the pan is displayed with the stability indicator, the
balance is calibrated. Remove the mass from the pan.
When Linearity calibration is completed, the balance automatically exits the
menu and stores any changes you have made.
SPAN CALIBRATION
Scout
II
balances are calibrated before shipment, however calibration can be
affected by changes in location, temperature, or rough handling. Check the bal-
ance with the calibration mass supplied. If calibration is required, proceed as
follows: NOTE:
1200g and 6000g balances are not supplied with masses.
1. With the balance ON, press and hold Zero On until
CAL
is displayed
then release it.
-C-
is momentarily displayed followed by the value of the
mass which must be placed on the pan.
Do not disturb the balance while
-C-
is displayed.
2. Place the required mass on the pan and momentarily press Zero On.
3. When the weight on the pan is displayed with the stability indicator, the
balance is calibrated. Remove the mass from the pan. If Err1 and/or Err4
is displayed, perform linearity calibration.
NOTE: 1200g, 4000g and 6000g balances are not supplied with masses