beautypg.com

Calibration, Span calibration, Linearity calibration – Ohaus Scout II Electronic Balances User Manual

Page 9

background image

9

LINEARITY CALIBRATION

1. With the balance OFF, press Zero On until

MENU

is displayed.

2. Repeately press Mode Off until

Lin

is displayed, then press Zero On.

-C-

is momentarily displayed followed by the value of the mass which must

be placed on the pan.

Do not disturb the balance while

-C-

is displayed.

3. Place the displayed value of the first mass on the pan and momentarily

press Zero On.

-C-

is displayed, then the value of the second mass to be

placed on the pan is displayed.

4. Place the displayed value of the second mass on the pan and momentarily

press Zero On.

5. When the weight on the pan is displayed with the stability indicator, the

balance is calibrated. Remove the mass from the pan.

When Linearity calibration is completed, the balance automatically exits the

menu and stores any changes you have made.

SPAN CALIBRATION

Scout

II

balances are calibrated before shipment, however calibration can be

affected by changes in location, temperature, or rough handling. Check the bal-

ance with the calibration mass supplied. If calibration is required, proceed as

follows: NOTE:

1200g and 6000g balances are not supplied with masses.

1. With the balance ON, press and hold Zero On until

CAL

is displayed

then release it.

-C-

is momentarily displayed followed by the value of the

mass which must be placed on the pan.

Do not disturb the balance while

-C-

is displayed.

2. Place the required mass on the pan and momentarily press Zero On.

3. When the weight on the pan is displayed with the stability indicator, the

balance is calibrated. Remove the mass from the pan. If Err1 and/or Err4

is displayed, perform linearity calibration.