Parts information, Care and maintenance, Troubleshooting – Ohaus Scout II Electronic Balances User Manual
Page 24: Activating units and parts counting, Parts counting, Replacement parts, Accessories, Error codes
![background image](/manuals/257212/24/background.png)
10
11
12
9
PARTS INFORMATION
If you require replacement parts or would like to purchase accessories,
please call Ohaus Corporation toll-free at (800) 526-0659. An Ohaus Prod-
uct Parts Specialist will be available to help you.
Error Codes
Err 0 Invalid checksum data. The balance may need recalibration,
particularly linearity calibration. If the error persists after
recalibration, the balance must be serviced.
Err 1 Sample being weighed exceeds the capacity of the balance. If
error occurs when the sample is within the balance capacity, bal-
ance may be incorrectly calibrated. An underload such as the pan
off of the balance could also display Err1. Perform span calibration
(one mass required).
Err 4 Incorrect or no calibration mass used when performing calibration
procedure. Error will flash momentarily, then balance will use previ-
ous calibration data. If balance fails to calibrate, call the service
department or return for servicing.
Err 5 Value is too large to display; exceeds 99999 counts. Possibly
caused by an invalid parts counting sample.
Err 6 Parts counting error - sample weighs <1d. Balance shows error
then exits parts counting.
Err 9 Internal data error. Return balance for servicing.
NOTE:
When the balance displays Err1 and Err4, perform
linearity calibration. Linearity calibration requires two
masses. If the balance does not calibrate after
attempting linearity calibration, call Ohaus Service De-
partment at 973 377-9000 or (800) 526-0659.
CARE AND MAINTENANCE
To keep the balance operating properly, keep the housing and platform
clean. If necessary, a cloth dampened with a mild detergent may be
used. Keep calibration masses in a safe dry place. Unplug the AC
Adapter when not in use. For long term storage, remove the battery.
TROUBLESHOOTING
SYMPTOM
No Display.
BAT Indicator is on.
Incorrect weight
reading.
Calibration proce-
dures do not work.
Unable to display
weight in a particu-
lar weighing unit.
Balance won't store
selections made in
menu.
PROBABLE CAUSE
Power Adapter not
connected.
Battery is dead.
Battery is weak.
Balance out of cali-
bration.
Balance was not
rezeroed before
weighing.
Incorrect calibration
masses being used.
Weighing unit not ac-
tivated in menu.
END selection was
not used to exit
menu.
REMEDY
Connect AC Adapter.
Replace battery.
Replace battery.
Calibrate the bal-
ance.
Press ReZero On
with no weight on the
pan, then weigh item.
Use correct masses.
See error codes note.
Use Units menu to
set desired units ON
(see menu).
You must use END to
ACTIVATING UNITS AND PARTS COUNTING
1. When the balance displays
, press ReZero On to display
the grams indicator "g" with its current setting or
or
.
2. To change the displayed setting, press Mode Off.
3. To accept the displayed setting, press ReZero On. The display will
advance to the next weighing unit.
4. Repeat steps 2 and 3 for each weighing unit and parts counting.
When the last unit has been accepted, the display will again indicate
.
5. Repeatedly press Mode Off to change to another menu item or to exit,
press Mode Off until
is displayed, then press ReZero On.
PARTS COUNTING
Parts counting must first be activated in the Units Menu to use it. Refer
to Menu section to activate parts counting. The balance will count parts
based on the weight of a reference sample of 5, 10, 20 or 50 parts. For
optimum results, the parts should be uniform in weight.
1. Place a container on the pan and press ReZero On to tare it.
2. Press and hold ReZero On until
is displayed, then release it.
The display shows SEt XX, where XX is the last used counting sample.
To change the sample size, repeatedly press Mode Off and stop at
the desired sample number. Choices are 5, 10, 20 or 50 pieces.
3. Add the selected number of parts to the container, then press ReZero On.
The display shows the number of parts added.
4. Add parts as desired and read the quantity on the display.
5. To read the weight of the parts, press Mode Off to change to any of the
activated weighing units. You can return to parts counting at any time
by repeatedly pressing Mode Off until the parts counting indicator is
displayed. Sample weight is retained as long as the balance remains on.
REPLACEMENT PARTS
Part No.
AC Adapters:
100/120 V ac US Plug
90524-66
220 V ac Euro Plug
90524-63
240 V ac UK Plug
90524-64
240 V ac Australian Plug
90524-65
Pan for 200g Scout, 4.0" (10cm) dia.
77064-00
Pan for 400g and 600g Scout, 5 x 5.75" (12.7cm x 14.6 cm) 300007-010
Weigh Below Hook
5227-04
200g Calibration Mass (200g and 400g Scout)
51025-06
300g
Calibration Mass (600g Scout)
51035-05
ACCESSORIES
Part No.
Linearity Masses: 100g (200g Scout)
51015-05
200g (400g Scout)
51025-06
300g (600g Scout)
51035-05
Security Device
76288-01
Scoops:
Aluminum,
3.62 x 4.50 x 1.0”/
9.20 x 11.34 x 2.54 cm 4590-10
Black anodized, aluminum,
3.62 x 4.50 x 1.0”/
9.20 x 11.34 x 2.54 cm 4590-30
Aluminum,
1.5 x 2.00 x 0.43”/
3.81 x 5.08 x 1.11 cm 5076-00
Gold anodized aluminum,
2.25 x 3.00 x 0.75”/
5.71 x 7.62 x 1.90 cm 5077-00